Appendix B. Start up and Shut down Procedure (Non-Mandatory)  


(a) Start up.
(1) Turn off ultraviolet sterilizer if so equipped.
(2) Turn on all blowers and cabinet illumination lights.
(3) Allow 5 minutes of operation to purge system, check flow alarm system audio and visual alarm function if so equipped.
(4) Decontaminate readily accessible interior surfaces with a disinfectant appropriate for the agents or suspected agents present.
(b) Shut down.
(1) Decontaminate and remove all items from interior work area.
(2) Decontaminate readily accessible interior surfaces with a disinfectant appropriate for the agents or suspected agents present.
(3) Turn on ultraviolet sterilizer if so equipped.
(4) Allow 5 minutes of operation to purge system.
(5) Turn off cabinet blower.
HISTORY
1. New Appendix B filed 10-11-94; operative 11-10-94 (Register 94, No. 41).